DSC and the glass transition: a polymer fingerprint that also predicts cold performance

Differential scanning calorimetry finds the glass transition temperature of a rubber. It supports polymer identification and is the single most useful number for HPLT duty.

Intermediate2 min readArticleUpdated 2026-10-23FKMHNBRFEPMVMQEPDM
Cover image: "DSC and the glass transition: a polymer fingerprint that also predicts cold performance", with a cold snowflake disc.

Rubber is a rubbery solid only above a temperature called the glass transition temperature (Tg). Below Tg the polymer chains are frozen and the part behaves like glass: stiff, brittle, no longer sealing. Differential scanning calorimetry (DSC) measures Tg by tracking the heat needed to warm a small sample.

Typical Tg ranges

Polymer Approximate Tg
Silicone (VMQ) about −120 °C
Polybutadiene (BR) about −100 °C
Natural rubber about −65 °C
EPDM about −50 to −55 °C
SBR about −50 to −60 °C, rising with styrene
Polychloroprene about −45 °C
NBR and HNBR roughly −45 to −15 °C, rising with acrylonitrile content
Acrylate rubber (ACM) roughly −40 to −15 °C depending on grade
FKM (standard VDF/HFP) about −20 to −15 °C
Low-temperature FKM about −40 °C and below, with special monomers
FEPM (TFE/propylene) close to 0 °C

These are indicative. Grade and composition move the number by several degrees.

What DSC adds to identification

Tg is a quick fingerprint. A nitrile or HNBR with Tg near −35 °C is probably a mid-ACN grade, and a result close to −15 °C points to a high-ACN type. A fluoroelastomer with a Tg around −17 °C is a standard type, while one near −35 °C is a low-temperature grade. When the IR spectrum says "fluoropolymer" and DSC says "−40 °C", you know to look for a low-temperature or perfluoroether-containing family.

Plasticisers lower the apparent Tg. A compound with 15 phr of plasticiser will show a lower Tg than the raw polymer, so measure both the cured part and the extracted rubber.

Why HPLT needs more than Tg

A low Tg is necessary but not sufficient. Seals fail in the cold because of stiffening, loss of recovery and low-temperature compression set, which depend on the network and fillers as well as the polymer. Tg should always be followed by tests such as TR10 and TR70 retraction and low-temperature compression set. Pascalor LT and ULT grades are specified on those, not on Tg alone.

Running DSC well

  • Use a small, representative sample, about 5 to 15 mg.
  • Run a heating cycle that erases thermal history, then a controlled second heating, and report the midpoint of the transition from the second run.
  • Quote the heating rate. A different rate shifts the number.

Sources and further reading

  • ASTM D3418 and ISO 11357-2, determination of glass transition temperature by DSC.
About this article. It is original writing prepared from public sources, standards and general industry practice. It is not a copy of any book or standard and it is not engineering advice. Confirm values against the current edition of the standard and test in your own application. Only analyse material you own or are entitled to analyse.